The blog

How to HPP270 vH2O2 probe is connected to the Indigo500 data processing unit

Data record vaporizing trend of hydrogen peroxide
Application engineer Nevon Mansour
Nevon Mansour
Application engineer
Published: Two month 22, 2023
Industrial manufacturing and process
Industrial measurement
Life science

In this video, visa application engineer Nevon Mansour demonstrates how to connect and configuration is suitable for the vaporization of hydrogen peroxide biological purification applicationsVisa HPP272 probe.Visa,Indigo500 series data processing unitIs suitable for visaIndigo compatible probeHost devices.Indigo500 series including Indigo520 and Indigo510 data processing unit.In the video below, Indigo520 Nevon used for demonstration.You can find the transcript at the bottom of the video.

Transcript:
Hello, everyone, my name is Nevon Mansour in visa application engineer.Today we will demonstrate how to use Indigo520 perform data records.We will be carried out in accordance with the will in the isolator vH2O2 as to set the HPP272 probe measurement.We have a Indigo520, HPP272 and temperature sensorTMP1.

From this screen, you can see the HPP272 will measure hydrogen peroxide concentration, relative saturation and TMP1 temperature of steam.Using the temperature sensor, you can from another location measurement temperature.1.13.1 and older versionIndigo520 provide data record function.(Check the version history).In order to realize data record, please check your data unit version is in compliance with the above requirements.

Please note that I have created the Web user interface.Before starting use PC to access the data processing unit, please do this, it is very important.If you need to update software or create a Web interface to the user's guide, please refer toIndigo520 user guideAnd we makeVideo data processing unit configuration.

I will be in the data processing unit and the establishment of physical connection between my PC.In order to do this, you need to network cable and Allen wrench.With an Allen wrench open data processing unit.

I connect the cable to the data processing unit and my PC.The last item on the operation of the configuration data processing unit is activated Web interface - find marked "press to activate Web interface" button, and then use the pen pointed items such as press the button.If you see the button lights up above the green light, then explain Web interface has been activated.

Now, we will be the IP address of the PC is configured to the same as the data processing unit.Open your network connection.Double click on the right side of the Internet port and select Properties (attributes).Select Internet protocol version 4 (Internet protocol version 4).Select the radio button: "Use the following IP address" (Use the following IP address).

You need to fill in the IP address, the address is 192.168.5.1.Then click OK (sure).Next, open a Web browser, type in the address (URL) field: https://192.168.5.20, followed by a colon and 8443 in English.Click "Advanced" (Advanced) and continue.Then you need to log in.

You can now see the measurement result, and can be in the top right corner of the "export data" button to export data.Choose the record type you want to export, such as 10 minutes, 1 hour, 6 hours, etc.Here we use for 10 minutes.Export data may take some time.

Now, after downloading data, you can analyze the data.You can see some basic information on the top, below are two probe data.So you first see is HPP272.Then we can scroll TMP1 temperature sensor data.

Hope this video will be helpful to you!Thank you for watching!

If you are using HPP270 series probe or Indigo data processing unit for biological purification have any questions, pleaseContact us.

VH2O2 biological purification: optimization of aseptic isolator cycle development

The webinars product manager by visa, Joni Partanen withArdien consultancy,Rick Nieskes chair together.Rick has more than 30 years of industry experience, and will show how to optimize drug isolator vH2O2Biological purification cycle development.Joni will be introducedVisa to pull the vH2O2The sensorTechnology, as well as to those exposed to the vH2O2Biological purification of ambient conditions of sensor calibration and maintenance is of great importance.